How to clean surgical instruments properly

Ultrasonic Cleaning - For micro and delicate instruments, use manual cleaning (step C). Instruments should be processed in a cleaner for the full recommended cycle time – usually 5 to 10 minutes. Place instruments in open position into the ultrasonic cleaner. Make sure that "Sharp" (scissors, knives osteotomies etc.) blades do not touch other instruments. All Instruments have to be fully submerged. Do not place dissimilar metals (stainless, copper, chrome plated, etc.) in the same cleaning cycle. Change solution frequently – at least as often as manufacturer recommends. Rinse instruments after ultrasonic cleaning with water to remove ultrasonic cleaning solution.